The new design model with opening water tank on the top.Strictly meet the standard of EN13060,which makes wash and cleaning much more convenient.

All-round computerized fuzzy control.Digital display modularized panel for easy operation.

Multi language menus are available. A special curve can displays all the sterilization process.

Mini printer can be attached to record the process of sterilization. It is installed with BOWICK&DICK which measures the penetration of water steam.

Automatic malfunction detection system makes maintenance more convenient.

   
 

Program Function

 

  1. EMERGENCY : 134°C/3min, One time vaccum, sterilization time 3 minutes, drying time 3 minutes.
  2. SOLID :
    • 134°C/4min, for nude and solid instrument. One time vaccum, sterilization time 4 minutes, drying time 5 minutes.

    • 121°C/20min, for nude and solid instrument. One time vaccum, sterilization time 20 minutes, drying time 5 minutes.
  3. HOLLOW :
    • 134°C/4min, packaged instrument or hollow hand pieces. Three times vaccum, sterilization time 4 minutes, drying time 8 minutes.

    • 121°C/20min, packaged instrument or hollow hand pieces. Three times vaccum, sterilization time 20 minutes, drying time 8 minutes.
  4. POROUS :
    • 134°C/4min,porous instrument and hand - pieces. Three times vaccum, sterilization time 4 minutes, drying time 10 minutes.

    • 121°C/20min, porous instrument and hand - pieces. Three times vaccum, sterilization time 20 minutes, drying time 10 minutes.
  5. SANGUINARY : 134°C/18min, sterilize the sanguinary or virulent bacillus on the instrument. Three times vaccum, sterilization time 18 minutes, drying time 18 minutes.
  6. RUBBER : 121°C/20min, rubber instrument. One time vaccum, sterilization time 20 minutes, drying time 5 minutes.
  7. GLASS : 121°C/20min, glass instrument. One time vaccum, sterilization time 20 minutes, drying time 4 minutes.
  8. TAMPON : 121°C/30min, tampon cotton. One time vaccum, sterilization time 30 minutes, drying time 15 minutes.
  9. SPECIAL : user - defined. The sterilization data by himself. user could set up the working temperature, vaccum times, srerilization time, drying time and etc. data before starting a cycle.
  10. HELIX/B&D : 134°C/3.5min, it is aHELIX and B&D test program. Three times vaccum, sterilization time 3.5 minutes, drying time 8 minutes.
  11. VACCUM : 5min, it is a vaccum test program.Vaccum time 5 minutes, keep pressure time 10 minutes.

 

Specification :

Model

Voltage/
Power

Chamber Size mm

Outside Size mm

Packing Size mm

Net.W kg

TANDA

II D 18

220 V/1500W

Φ245*320

435*606*392

500*633*437

47

TANDA

II D 23

220 V/1700W

Φ245*450

435*716*392

500*743*437

53

   
 